a. Compact and aesthetically pleasing in appearance, with a unique structural design;
b. High-resolution LCD display provides clear and beautiful visuals;
c. It could offer a cost-effective solution for the detection of trace moisture in high-end glove boxes;
d. Resistant to a small number of corrosive gases;
e. It has sensor over-limit protection and automatic recovery functions;
f. Menus are locked to prevent accidental changes to instrument parameters and affect its performance;
g. It is available for installation using KF40 flanges or threaded connections for in-situ measurement, or using a test chamber for ventilation measurement;
h. The circuit employs digital processing technology, isolation amplification, and band-pass design, featuring strong anti-interference capabilities;
i. RS485 bidirectional communication function enables direct communication with computers or other digital communication devices.
l Measurement Principle: Electrolyte
l Display Mode: LCD display with high resolution
l Measurement Range: 0~500ppmV H2O (range customizable)
l Measurement Accuracy:0~10ppm±3%FS,10~500ppm±5%FS
l Repeatability:±2% (when 10ppm)
l Stability:±2%FS/7d
l Resolution: 0.01PPm
l Response Time 63%: T63≤3min
l Output Port: 4-20mA (load resistance<1KΩ)
l Communication Mode: RS485
l Power Supply: DC24V±10%, 0.2A
l Storage Environment Humidity: <90%RH, non-condensing (When the storage and transportation environment temperature is lower than 5℃, the transmitter should be placed in a normal working temperature environment for more than 4 hours before use.)
l Operation Environment Humidity: <100%RH, non-condensing
l Sample Mode: Free diffusion (other method is customizable)
l Sample Gas Temperature: -10℃~+50℃
l Coating Cycle: 6 months (The higher the concentration measured, the shorter the coating cycle.)
l Calibration Cycle: 1 year (recommended)
l Sensor Service Life: >24 months (under normal conditions of use)
l Gas Path Connection: NPT 1/8 internal thread for calibration cover
l Installation Method: KF40 flange installation
It is used in glove boxes, research and production of lithium-ion batteries, inert atmosphere welding, scientific laboratories, pharmaceuticals, nuclear industry, and chemical applications.