Product Name: HT-TA261 Nitrogen/ Oxygen Transmitter
器特點
Instrument Features

a. The instrument is small and easy to maintain;

b. The imported new-type zirconia sensor has a long service life, high sensitivity and rapid response speed;

c. There are N2and O2 types of display mode, and N2 displays in (N2%=100-O2%);

d. Automatic temperature compensation could eliminate the impact of environment temperature changes on the measured value;

e. Upper and lower alarm limit can be set among the full range;

f. The instrument allows multi-point fractional liner calibration, meeting the need for an accurate measurement of gas concentration among the full range.

g. Built-in standard RS485 communication port allows a unidirectional or bidirectional communication with computer or other digital communication equipment.

術參數
Technical Parameters

l Measurement Principle: Zirconia

l Display Mode: 5 dight high-light LED display, external display module available (optional)

l Measurement Range: 10/100/1000PPm~21.000% O2/N2

l Measurement Accuracy:10PPm~21.00%≤±1%FS

          10PPm~1.00%≤±2%FS

          10PPm~100PPm≤±5%FS    

l Repeatability: ≤±1.0%FS

l Response Time: T90<30S

l External Port: 4-20mA.DC (Non-isolated output, load resistance<1KΩ) 

1 way of programmable alarm output (Dry contact type, maximum load 220VAC/2A)

Display module port

l Communication Mode: RS485

l Power Supply: DC24V±10%,1.5A

l Environment Temperature: -10℃~+50℃

l Storage Environment Humidity:<90%RH, non-condensing

l Operation Environment Humidity:<100%RH, non-condensing

l Sample Gas Temperature: 0℃~60℃

l Sampling Mode: Gas in

l Gas Inlet Pressure: Stable pressure (control the pressure within the recommended limits to make sure the flow is within 400~600mL/min)

l Gas Outlet Pressure: Free venting (under safe conditions)

l Background Gas: Mixed gas of nitrogen and oxygen

l Sensor Service Life: >5 years (under normal conditions of use)

l Gas Circuit Port: Φ6 plastic tube fitting

l Installation Method: Instrument: embedded mounting

Oxygen sensor: fixing clamp

用場合
Application Scenarioses

l Air separation industry: air separation fractionation, nitrogen generator, gas purification;

l Chemical/petrochemical industry: chemical production, high purity gas production;

l Electronics: reflow soldering, wave soldering, soldering powder production, semiconductor processing furnace;

l Metal: heat treatment/low temperature annealing, steel, alloy or powder metal production;

l Regular application: environmental monitoring, glove box;

l All kinds of purification equipment, chemical process, glove box, protective gas furnace and other online analysis of oxygen concentration measurement.

貨須知
Ordering Instructions

Measurement range

Measured gas pressure

Background gas composition

Contact Number

028-85044052/053

Top
網站地圖